Full Body MRI Magnet Design

  • Category: Computational Mechanics
  • Funding Institute: National Science Foundation
  • Project start date: August 2013

Full Body MRI Magnet Design

High temperature superconductors such as MgB2 focus on conduction cooling of electromagnets that eliminates the use of liquid helium. With the recent advances in the strain sustainability of MgB2, a full body 1.5 T conduction cooled magnetic resonance imaging (MRI) magnet shows promise. In this article, a 36 filament MgB2 superconducting wire is considered for a 1.5 T full-body MRI system and is analyzed in terms of strain development. In order to facilitate analysis, this composite wire is homogenized and the orthotropic wire material properties are employed to solve for strain development using a 2D-axisymmetric finite element analysis (FEA) model of the entire set of MRI magnet. The entire multiscale multiphysics analysis is considered from the wire to the magnet bundles addressing winding, cooling and electromagnetic excitation. The FEA solution is verified with proven analytical equations and acceptable agreement is reported. The results show a maximum mechanical strain development of 0.06% that is within the failure criteria of −0.6% to 0.4% (−0.3% to 0.2% for design) for the 36 filament MgB2 wire. Therefore, the study indicates the safe operation of the conduction cooled MgB2 based MRI magnet as far as strain development is concerned.